HP Plus Rack Nitrogen Generator
The VICI DBS® HP Plus Rack produces nitrogen by utilizing a combination of filtration and pressure swing adsorption (PSA) technology.
Overview
Zero grade nitrogen in a 19” rack.
Standard compressed air is filtered by high efficiency coalescing filters to remove all contaminants down to 5 micron. For ultra sensitive applications the 19” rack units also include the additional heated catalyst module to ensure hydrocarbons are removed to <0.1 ppm. The air then passes through two columns filled with a carbon molecular sieve (CMS) which removes O2, CO2 and moisture. These are desorbed to atmosphere during the pressure swing cycle leaving a supply of ultra pure nitrogen.
Features & benefits
What sets the HP Plus Rack apart.
Features
- Produces a continuous supply of zero grade nitrogen
- On-demand supply 24/7
- Flow rate: up to 500 mL/min
- Purity: +99.999% & <0.1 ppm hydrocarbons
- Pressure: up to 5 barg (75 psig)
- Proprietary carbon molecular sieve technology
- Easy to install, operate and maintain
Benefits
- Eliminates dangerous high pressure cylinders helping to keep your employees safer
- Removes the logistics, inconvenience, downtime and costs of a cylinder system
- Flow capacity to match your specific instrument demands
- Ideal for analyzer applications – stable baseline with increased sensitivity and repeatability
- Exceeds the requirements for the most demanding applications
- Superior nitrogen purification with long life catalyst
- Peace of mind
- Improve your laboratory work flow and productivity
Applications
What you can use the HP Plus Rack for.
Specifications
The HP Plus Rack in detail.
| Models & specs | HP Plus Rack 500 |
|---|---|
| Flow mL/min | 500 |
| Purity | +99.999% |
| Hydrocarbon purity (measured as methane) | <0.1 ppm |
| Dewpoint | -50 °C |
| Outlet pressure barg (psig) | up to 5 (75) |
| Inlet pressure barg (psig) | 7 to 10 (100 to 160) |
| Actual inlet air requirement litres – at 8 barg (116 psig) | 11 |
| Recommended compressor air inlet – at 8 barg (116 psig) | 22 |
| Pressure drop barg (psig) | 1.5 (22) |
| Inlet air quality | Clean dry compressed air ISO 8573-1:2010 Class 1.2.1 |
| Technology | Carbon molecular sieve |
| Warm up time (minutes) | 60 |
| LED indicators | Power on/off, system ready, errors |
| Electrical supply | 110–120 V 60 Hz / 220–240 V 50 Hz |
| Power consumption (watts) | 270 |
| Noise level | minimal |
| Dimensions mm (in) | 19”W x 3UH x 500D (19”W x 3UH x 19.6D) |
| Weight kg (lbs) | 10 (22) |
| Shipping dimensions mm (in) | 720W x 565H x 375D (28.3W x 22.2H x 14.7D) |
| Shipping weight kg (lbs) | 14 (31) |
| Operating temp °C (°F) | 15 to 35 (59 to 95) |
| Inlet connection | 1/4” OD Swagelok or 6 mm OD rapid fitting |