GC Plus Rack Zero Air Generator
The VICI DBS® GC Plus Rack Zero Air Generator utilizes compressed air that is prefiltered to 5 microns and then purified using a state of the art combined heated catalyst module. The output zero grade air is free from total hydrocarbons to <0.1 ppm, making it ideal for all FID applications. This purity level (measured as methane) produces a low signal to noise ratio, ensuring a flat and stable analyzer baseline.
Overview
Zero air in 19-inch format – for your analyzers.
With no moving parts and silent operation, the generator is extremely reliable and can be installed directly in the laboratory. With flow rates up to 15 L/min one system can support more than 30 FIDs. With short pay back time, minimal maintenance and operator attention they are an ideal addition to any analyzer application.
Features & benefits
What sets the GC Plus Rack apart.
Features
- Produces a continuous supply of zero grade air
- On-demand supply 24/7
- Flow rate: 1 to 15 L/min
- Purity: <0.1 ppm THC
- Pressure: 4.5 to 10 barg (65 to 145 psig)
- Proprietary platinum catalyst technology
- Easy to install, operate and maintain
Benefits
- Eliminates dangerous high pressure cylinders helping to keep your employees safer
- Removes the logistics, inconvenience, downtime and costs of a cylinder system
- Ideal for all destructive GC detector applications – stable baseline with increased sensitivity and repeatability
- Ideal for all GC detector applications – stable baseline with increased sensitivity and repeatability
- Meets and exceeds the requirements for the most demanding analyzers
- Superior air purification with long life catalyst technology
- Peace of mind
- Improve your laboratory work flow and productivity
Applications
What you can use the GC Plus Rack for.
Analyzer applications
- Online GC-FID
- Mud logging
- Total Hydrocarbon Analyzer (THA) detector gas
Technical specifications
The GC Plus Rack in detail.
| Models & specs | GC Plus 1800 Rack | GC Plus 5000 Rack | GC Plus 10000 Rack | GC Plus 15000 Rack |
|---|---|---|---|---|
| Flow mL/min | 1800 | 5000 | 10000 | 15000 |
| Purity – hydrocarbons | <0.1 ppm | |||
| Purity – CO | <0.1 ppm | |||
| Inlet pressure barg (psig) | 4.5 to 10 (65 to 145) | |||
| Inlet air quality | Clean dry compressed air ISO 8573-1:2010 Class 1.2.1 | |||
| Max outlet pressure barg (psig) | 1 barg (15 psig) drop from inlet | |||
| Max HC in | 100 ppm | |||
| Max CO in | 50 ppm | |||
| Technology | Platinum catalyst | |||
| LED indicators | Power on/off, system ready, errors | |||
| Warm up time (minutes) | 45 | |||
| Electrical supply | 110–120 V 60 Hz / 220–240 V 50 Hz | |||
| Power consumption (watts) | 200 | 550 | 550 | 550 |
| Dimensions mm (inches) | 19" W × 3U H × 500 D (19" W × 3U H × 19.6 D) | |||
| Weight kg (lbs) | 15 (33) | |||
| Shipping dimensions mm (inches) | 720 W × 565 H × 375 D (28.3 W × 22.2 H × 14.7 D) | |||
| Shipping weight kg (lbs) | 19 (42) | |||
| Operating temp °C (°F) | 15 to 35 (59 to 95) | |||
| Inlet connection | 1/4" compression | |||
| Outlet connection | 1/8" compression | |||
| Certification | CE, FCC, MET (UL and CSA compliant) | |||